With over 20,000 sold, Explorer sets the standard for data logging in Primary, Middle and Junior education.
With built in sensors and software supplied it is ready to use straight from the pack.
Single Explorer packs start at UK £169 + vat or save upto £200 with Explorer multi packs.
Support
Q:
I have fitted four batteries but nothing happens on the display when I press any of the Explorer buttons?A:
Remove and check that all batteries are fitted correctly as shown in the battery holder. They must all be new Alkaline batteries - others will not work properly and must not be used.Q:
I have other USB accessories for my computer which take their power from the computer but my USB Explorer does not seem to work without batteries?A:
Because Explorer needs to be able to work away from the computer to log and store sensor data, you always need to have batteries installed.Q:
Can I turn off internal sensors I am not using?A:
Most Explorer software lets you switch on and off sensor traces. If you want to switch off an internal sensor you can plug in a standard unused 3.5mm stereo jack plug (from most electrical suppliers) which will stop that sensor working.Q:
Sometimes when I am running a long experiment with Explorer graphing live on my computer it seems to stop logging.A:
Some computer 'screen savers' can interrupt continuous datalogging, so we suggest screen savers are turned off.Q:
My Explorer seems to be doing some odd things but my batteries are OK - can I reset Explorer?A:
Occasionally any microprocessor based system may need resetting. To fully reset Explorer, remove 1 battery, wait for around 10 seconds and then while holding the Blue button down refit the battery - Explorer should now fully reset.Q:
I have tried using the fibre optic to enable me to measure light level but I find the readings are quite a bit lower.A:
Fibre optics do lose some light but you can reduce the loss by carefully cutting the fibre square with a sharp knife.Q:
I wish to use Explorer as a hand held meter but the display automatically turns off after a while - can I keep it on?A:
Keep the meter display on by pressing and HOLDING the green button on for more than 4 seconds when switching on Explorer.Q:
Can I use Explorer with Windows Vista?A:
Yes. Provided you run LogIT Lab version 3.21n or better and Datalogging/Junior Insight 4.21 or better.